10 Operation
▪ Setting the ultrasonic electronics for configuration
146
1
Press calibration switch (1) upwards to open it.
▪ Screw down the cover of the ultrasonic electronics
1
Each time you open the ultrasonic electronic you have to expect that the O-
ring is damaged. Therefore, this damaged O-ring has to be replaced with a
new one in general. (RMG provides a replacement kit with O-ring, grease, ...
sales number: 38.03.001.00)
2
Screw the cover on with both hands.
3
Insert the special tools into the boreholes.
4
Tighten the cover with the special tools.
5
Screw in threaded pin hand tight.
6
Switch on the system.
Manual USM GT400 · EN10 · May 2021
1
Calibration switch
Fig. 10.4: Opening the calibration switch
1
Cover
2
Position O-ring
Fig. 10.5: Screw down the cover