Testing Performance
To test the single-clock, multiple-edge, state acquisition (logic analyzer)

Connect the logic analyzer

Using the 6-by-2 test connectors, connect the first combination of logic analyzer
1
clock and data channels listed in one of the following tables to the pulse generator.
If you are testing an Agilent 1670G or Agilent 1671G, you will repeat this test for the second
combination.
Using the SMA cables, connect channel 1, channel 2, and trigger from the
2
oscilloscope to the pulse generator.
Connect the Agilent 1670G or Agilent 1671G Logic Analyzer to the Pulse Generator
Testing
Connect to
Combinations
8133A
Channel 2 Output
1
Pod 1, channel 3
Pod 3, channel 3
Pod 5, channel 3
Pod 7, channel 3
2
Pod 1, channel 11
Pod 3, channel 11
Pod 5, channel 11
Pod 7, channel 11
3–44
Connect to
Connect to
8133A Channel 2
8133A Channel 1
Output
Output
Pod 2, channel 3
J-clock
Pod 4, channel 3
Pod 6, channel 3
Pod 8, channel 3 *
Pod 2, channel 11
J-clock
Pod 4, channel 11
Pod 6, channel 11
Pod 8, channel 11 *
*Agilent 1670G only