Testing 16850 Series Performance

Connect the Test Equipment

Connect the Logic Analyzer Pod to the Pulse Generator

N O T E
32
Connect one U4203A Flying Lead Probe Set to Pods 1/2 of
1
the 16850 series logic analyzer.
Connect the Pod 1 U4203A Flying Lead Probe Set's CLK
2
lead to the pin strip of the SMA/Flying Lead connector at
the pulse generator's Channel 1 OUTPUT.
ground clip
Be sure to use the black ground clip (supplied with the U4203A Flying Lead
Probe Set) and orient the leads so that the black clip is connected to one of
the SMA/Flying Lead connector's ground pins!
3
Connect the Pod 1 U4203A Flying Lead Probe Set's CLK
lead to the SMA/Flying Lead connector at the pulse
generator's Channel 1 Output. Again, be sure to use the
black ground clip and orient the leads so that the black
clip is connected to ground.
Connect the Pod 1 U4203A Flying Lead Probe Set's bits 2
4
and 10 to the SMA/Flying Lead test connector's pin strip
connector at the pulse generator's Channel 2 OUTPUT.
Connect the Pod 1 U4203A Flying Lead Probe Set's bits 6
5
and 14 to the SMA/Flying Lead test connector's pin strip
connector at the pulse generator's Channel 2 Output.
16850 Series Portable Logic Analyzer Service Guide