Testing Performance
This chapter tells you how to test the performance of the logic analyzer against the
specifications listed in chapter 1. To ensure the logic analyzer is operating as
specified, you perform software tests (self-tests) and manual performance tests on
the analyzer. The logic analyzer is considered performance-verified if all of the
software tests and manual performance tests have passed. The procedures in this
chapter indicate what constitutes a "Pass" status for each of the tests.
The Logic Analyzer Interface
To select a field on the logic analyzer screen, use the arrow keys to highlight the field,
then press the Select key. For more information about the logic analyzer interface,
refer to the Agilent 1670G-Series Logic Analyzers User's Guide.
Test Strategy
For a complete test, start at the beginning with the software tests and continue
through to the end of the chapter. For an individual test, follow the procedure in the
test. The examples in this chapter were performed using an Agilent 1670G. Other
analyzers in the series will have appropriate pods showing on the screen.
With the oscilloscope option, ensure that the operational accuracy calibration has
been done before doing the performance verification tests (see chapter 4).
The performance verification procedures starting on page 3–8 are each shown from
power-up. To exactly duplicate the setups in the tests, save the power-up
configuration to a file on a disk, then load that file at the start of each test.
If a test fails, check the test equipment setup, check the connections, and verify
adequate grounding. If a test still fails, the most probable cause of failure would be
the acquisition board.
Test Interval
Test the performance of the logic analyzer against specifications at two-year intervals.
Performance Test Record
A performance test record for recording the results of each procedure is located at
the end of this chapter. Use the performance test record to gauge the performance
of the logic analyzer over time.
Test Equipment
Each procedure lists the recommended test equipment. You can use equipment that
satisfies the specifications given. However, the procedures are based on using the
recommended model or part number. Before testing the performance of the logic
analyzer, warm-up the instrument and the test equipment for 30 minutes.