Advertisement

General Information

Specifications (oscilloscope)

Specifications (oscilloscope)

The Agilent 1670G Logic Analyzers with an oscilloscope also include the following
specifications:
(*,1)
Bandwidth
Time Interval Measurement Accuracy
(*)
DC Offset Accuracy
Voltage Measurement Accuracy
(*)
Trigger Sensitivity
(*)
Input R (selectable)
* Specifications (valid within ± 10° C of auto-calibration temperature, excluding bandwidth−see note 1 for bandwidth specification.)
1. Upper bandwidth reduces by 2.5 MHz for every degree C above 35° C.
2. Specification applies to the maximum sampling rate. At lower rates, the specification is:± [(0.005% x ∆t)+(2 x 10
(0.15 x sample interval)] for bandwidth limited signals (t
3. Digitizing level = (#vertical divisions)(

Specifications (pattern generator)

There are no specifications for the pattern generator option of the Agilent 1670G-Series Logic
Analyzers.
1–4
DC to 500 MHz (real time, dc-coupled)
(*, 2)
±[(0.005% X ∆t)+
(2 x 10
±(1.0% of channel offset + 2.0% of full scale)
(*, 3)
± [(1.25% of full scale + offset accuracy)
+ (0.016 x V/div)]
DC to 50 MHz: 0.063 x full scale,
50 to 500 MHz: 0.125 x full scale
1 MΩ: ±1%, 50 Ω: ±1%
= 1.4 x sample interval). Sample interval is defined as
r
1
1
# bits in ADC
)(
), where LSB=2
2
LSB
−6
x delay setting)+150 ps]
-6
x delay setting) +
1
sample rate