Testing Performance
To test the single-clock, single-edge, state acquisition (logic analyzer)
Set up the oscilloscope.
3
a Select Setup, then select Default Setup.
b Configure the oscilloscope according to the following table.
Oscilloscope Setup
Acquisition
Display
Averaging: On
Graticule
# of averages: 16
Graphs: 2
Channel 1
Alternate Scale
Attenuation: 20.00:1
Scale: 200 mV/div
Offset: −1.300 V

Set up the logic analyzer

Set up the Configuration menu.
1
a Press the Config key.
b In the Configuration menu, assign all pods to Machine 1. To assign the pods, select
the pod fields, then select Machine 1 in the pop-up menu.
c Select the Type field in the Analyzer 1 box, then select State Compare.
3–18
Trigger
Level: -250 mV
Channel 2
Define meas
Alternate Scale
Thresholds: user-defined
Attenuation: 20.00:1
Scale: 200 mV/div
Offset: -1.300 V
[Shift] ∆ Time
Stop src: channel 2 [Enter]
Units: Volts
Upper: -980 mV
Middle: -1.30 V
Lower: -1.62 V