Calibrating and Adjusting
To test the CAL OUTPUT ports
Set up the logic analyzer
Set up the Calibration menu.
1
a Press the Waveform key.
b Press the Waveform key again. At the pop up, select Scope Calibration.
c Select the Mode field, then select Service Cal.
d Select the Procedure field, then select Cal BNC Out.
e Select Signal Probe Comp, then select DC Level.
f Select the DC volts field, and set it to 0 V.
4–7