Test The Next Channels - Agilent Technologies 1670G Series Service Manual

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

Testing Performance
To test the multiple-clock, multiple-edge, state acquisition (logic analyzer)
Press the blue shift key, then press the Run key. If two to four acquisitions are
10
obtained without the "Stop Condition Satisfied" message appearing, then the test
passes. Press Stop to halt the acquisition. Record the Pass or Fail results in the
performance test record.
Test the next clocks.
11
a In the logic analyzer Format menu, select Master Clock.
b Turn off and disconnect the clocks just tested.
c Repeat steps 2 through 11 for the next clock edges listed in the table in step 4, until all
listed clock edges have been tested.
Test the next setup/hold combination.
12
a In the logic analyzer Format menu, select Master Clock.
b Turn off and disconnect the clocks just tested.
c Repeat steps 1 through 12 for the next setup/hold combination listed in step 1 on
page 3-36, until all listed setup/hold combinations have been tested.
When aligning the data and clock waveforms using the oscilloscope, align the waveforms
according to the setup time of the setup/hold combination being tested, +0.0 ps or −100 ps.

Test the next channels

Connect the next combination of data channels and clock channels, then test them.
Start on page 3–32 "Connect the logic analyzer," connect the next combination, then continue
through the complete test.
3–40