Testing Performance
To test the threshold accuracy (logic analyzer)

Set up the logic analyzer

Press the Config key.
1
Unassign Pods 3 and 4, Pods 5 and 6, and Pods 7 and 8. To unassign the pods, select
2
the pod field. In the pop-up menu, select Unassigned.
Set up the Format menu.
3
a Press the Format key.
b Select the field to the right of Pod A1, then select ECL in the pop-up menu.
c Move the cursor over the Pod A1 channel selection field. Press the Clear Entry button
until all channels are assigned (all '*'), then press Done.
d Move the cursor over the Data on Clocks channel selection field, then press Select.
Rotate the RPG knob to move the channel selector to the channel indicated in the
following table, then press Select. Press Done.
Pod
Channel
A1, A3, A5, A7
0
A2, A4, A6, A8
1
3–10