Performance Test Record (oscilloscope)
Test
Settings
Voltage
Measurement
Accuracy
Zero Input
Channel 1
Zero Input
Channel 2
Zero-input offset
Offset Accuracy
4 V/Div
Channel 1
1 V/Div
100 mV/Div
10 mV/Div
4 V/Div
Channel 2
1 V/Div
100 mV/Div
10 mV/Div
DC-input offset
-35.00 V
Channel 1
-10.00 V
-2.00 V
+2.00 V
+10.00 V
+35.00 V
-35.00 V
Channel 2
-10.00 V
-2.00 V
+2.00 V
+10.00 V
+35.00 V
Performance Test Record (oscilloscope)
Results
Limits
-13.7 V to -14.3 V
-3.43 V to -3.57 V
-1.37 V to -1.43 V
-137.0 mV to -143.0 mV
143.0 mV to 137.0 mV
1.43 V to 1.37 V
3.57 V to 3.43 V
14.3 V to 13.7 V
-13.7 V to -14.3 V
-3.43 V to -3.57 V
-1.37 V to -1.43 V
-137.0 mV to -143.0 mV
143.0 mV to 137.0 mV
1.43 V to 1.37 V
3.57 V to 3.43 V
14.3 V to 13.7 V
0.00 V±320.0 mV
0.00 V±80.0 mV
0.00 V±8.0 mV
0.00 V±800.0 µV
0.00 V±320.0 mV
0.00 V±80.0 mV
0.00 V±8.0 mV
0.00 V±800.0 µV
-35.4 V to -34.6 V
-10.1 V to -9.90 V
-2.02 V to -1.98 V
+1.98 V to +2.02 V
+9.90 V to +10.1 V
+34.6 V to +35.4 V
-35.4 V to -34.6 V
-10.1 V to -9.90 V
-2.02 V to -1.98 V
+1.98 V to +2.02 V
+9.90 V to +10.1 V
+34.6 V to +35.4 V
Testing Performance
Measured
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
3–89