Agilent Technologies 1670G Series Service Manual page 63

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

Using the Delay mode of the pulse generator channel 1, position the pulses
8
according to the setup/hold combination selected, +0.0 ps or -100 ps.
a On the Oscilloscope, select [Define meas] Define ∆ Time - Stop edge: falling.
b On the oscilloscope, select [Shift] ∆ Time. Select Start src: channel 1, then select
[Enter] to display the setup time (∆ Time(1)-(2)).
c Adjust the pulse generator channel 1 Delay until the pulses are aligned according the
the setup time of the setup/hold combination selected, +0.0 ps or -100 ps.
Disregard the oscilloscope Period(2) value. The settings provided in this procedure
measure the period from rising edge to rising edge, which is not a valid measurement.
Select the clocks to be tested.
9
a Select the clock field to be tested, then select the clock edges as indicated in the table.
Clocks
J↓ + K↓ + L↓ + M↓
b Select Done to exit the Master Clock menu.
To test the multiple-clock, multiple-edge, state acquisition (logic analyzer)
Testing Performance
3–39