Test the ECL Threshold
Set up the Format menu.
1
a Press the Format key.
b Select the field to the right of Pod A1, then select ECL in the pop-up menu.
Test the high-to-low transition.
2
a On the DC power source, enter -1.438 V.
b On the logic analyzer, press Run. The display should show all channels at a logic "0"
Testing Performance
To test the threshold accuracy (logic analyzer)
3–13