Calibrating and Adjusting
To test the CAL OUTPUT ports
To test the CAL OUTPUT ports
This test does not need to be performed every time an operational accuracy
calibration is done. Testing the oscilloscope CAL OUTPUT ports checks the following:
DC CAL OUTPUT voltage
AC CAL OUTPUT voltage
This test verifies that the CAL OUTPUT voltages are operating within limits, so that
they can provide accurate calibration for the oscilloscope operational accuracy
calibration and probe calibration.
Equipment Required
Equipment
Critical Specifications
Digital Multimeter
0.1 mV resolution, better than 0.005% accuracy
Cable
BNC (m)(m) 48-inch
Adapter
BNC (f) to Dual Banana Plug
Set up the equipment
Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes if you have not already done so.
4–6
Recommended
Model/Part
3458A
8120-1840
1251-2277