To perform the self-tests

The self-tests verify the correct operation of the logic analyzer. Self-tests can be
performed all at once or one at a time. While testing the performance of the logic
analyzer, run the self-tests all at once.
The performance verification (PV) self-tests consist of system PV tests and analyzer
PV tests. For the oscilloscope option there are oscilloscope PV tests, and for the
pattern generator option there is a pattern generator test.
These procedures assume the files on the PV disk have been copied to the /SYSTEM
subdirectory on the hard disk drive. If they have not already been copied, insert the
PV disk in the flexible disk drive before starting this procedure.
Disconnect all inputs, then turn on the power switch. Wait until the power-up tests
1
are complete.
Press the System key. Select External I/O, then select Test in the pop-up menu.
2
Select the box labeled Load Test System.
3
Select the Analy PV field. then select SysPV in the pop-up menu. Select the field
4
next to Sys PV, then select System Test in the pop-up menu.