Connect And Configure The Logic Analyzer To Test Other Pods - Agilent Technologies 16900 Series Service Manual

Logic analyzer modules
Hide thumbs Also See for 16900 Series:
Table of Contents

Advertisement

1 Disconnect all flying lead probes from the 2 x 9 test fixtures.
2 Connect the second E5383A Flying Lead Probe set to the next pod to be
NOTE:
3 Connect the even bits of Pod 2 to the 2 x 9 test fixture at the pulse
4 Connect the odd bits of Pod 2 to the 2 x 9 test fixture at the pulse
5 If the pod to be tested (Pod 2 in this example) has a clock bit, connect it to
NOTE:
6 Connect the clock bit from Pod 1 to the 2 x 9 test fixture at the pulse
7 In the Analyzer Setup window, Buses/Signals tab, un-assign all channels of
8 Assign all channels of the pod to be tested. If the pod to be tested has a
9 Select the Sampling tab.
10 Select the Sampling Positions button.
Chapter 5: Troubleshooting
To test the cables

Connect and configure the logic analyzer to test other pods

tested.
In this example the pod to be tested will be Pod 2.
generator's OUTPUT 2.
generator's OUTPUT 2.
the 2 x 9 test fixture at the pulse generator's OUTPUT 2.
Pods 5 and 6 of the 16910A logic analyzer do not have clock bits. All others
do.
generator's OUTPUT 2.
Pod 1.
clock, then assign it too (place a check mark in the box under the channel
name).
80

Advertisement

Table of Contents
loading

This manual is also suitable for:

16910a16911a