Set up the Format menu.
2
a Press the Format key.
b Select the field to the right of each Pod field, then select ECL. The screen does not
show all Pod fields at one time. Use the knob to access more Pod fields.
Set up the Trigger menu.
3
a Press the Trigger key. Select Modify Trigger, then select Clear Trigger, then select All
in the pop-up menu.
b Select Count Off. Press Select again, then select Time in the pop-up menu. Select
Done to exit the menu.
c Select Acquisition Control. If the Acquisition Control is "Manual," use the cursor keys
to move the cursor to the Acquisition Mode field. Press the "Select" key to toggle this
field to "Automatic."
d In the Acquisition Control menu, use the knob to set the Memory Length to 4096.
Press the Done key.
e Select the field labeled 1 under the State Sequence Levels. Select the field labeled
"anystate", then select "no state." Select Done to exit the State Sequence Levels menu.
To test the multiple-clock, multiple-edge, state acquisition (logic analyzer)
Testing Performance
3–31