Agilent Technologies 1670G Series Service Manual page 222

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

Theory of Operation
RS-232-C
RS-232-C
The logic analyzer interfaces with RS-232-C communication lines through a standard
25-pin D connector. The logic analyzer is compatible with RS-232-C protocol. When
a hardwire handshake method is used, the Data Terminal Ready (DTR) line, pin 20
on the connector, is used to signal if space is available for more data in the logical I/O
buffer. Pin outs of the RS-232-C connectors are listed in the following table.
RS-232-C Signal Definitions
Pin
Function
Number
1
Protective Ground
2
Transmitted Data (TD)
3
Received Data (RD)
4
Request to Send (RTS)
5
Clear to Send (CTS)
6
Data Set Ready (DSR)
7
Signal Ground
8
Data Carrier Detect (DCD)
20
Data Terminal Ready (DTR)
23
Data Signal Rate Selector
8–32
RS-232-C
Signal Direction and Level
Standard
AA
Not applicable
BA
Data from Mainframe
High = Space = "0" = +12 V
Low = Mark = "1" = –12 V
BB
Data to Mainframe
High = Space = "0" = +3 V to +25 V
Low = Mark = "1" = –3 V to –25 V
CA
Signal from Mainframe
High = ON = +12 V
Low = OFF = –12 V
CB
Signal to Mainframe
High = ON = +3 V to +12 V
Low = OFF = –3 V to –25 V
CC
Signal to Mainframe
High = ON = +3 V to +25 V
Low = OFF = –3 V to –25 V
AB
Not applicable
CF
Signal to Mainframe
High = ON = +3 V to +25 V
Low = OFF = –3 V to –25 V
CD
Signal from Mainframe
High = ON = +12 V
Low = OFF = –12 V
CH/CI
Signal from Mainframe
Always High = ON = +12 V