Agilent Technologies 1670G Series Service Manual page 45

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

Connect the Agilent 1672G or Agilent 1673G Logic Analyzer to the Pulse Generator
Testing
Connect to
Combination
8133A
Channel 2 Output
1
Pod 1, channel 3
Pod 2, channel 3
Pod 3, channel 3
Pod 4, channel 3
Activate the data channels that are connected according to one of the previous
3
tables.
a Press the Format key.
b Select the field showing the channel assignments for one of the pods being tested,
then press the Clear entry key. Using the arrow keys, move the selector to the data
channels to be tested, then press the Select key. An asterisk means that a channel is
turned on. When all the correct channels of the pod are turned on, press the Done
key. Follow this step for the remaining pods.
To test the single-clock, single-edge, state acquisition (logic analyzer)
Connect to
8133A Channel 2
Output
Pod 1, channel 11
Pod 2, channel 11
Pod 3, channel 11*
Pod 4, channel 11*
*Agilent 1672G only
Testing Performance
Connect to
8133A Channel 1
Output
J-clock
3–21