c On the oscilloscope, select [Shift] ∆ Time, then select [Enter] to display the setup time
(∆ Time(1)-(2)).
d Adjust the pulse generator channel 1 Delay until the pulses are aligned according the
the setup time of the setup/hold combination selected, +0.0 ps or -100 ps.
Disregard the oscilloscope Period(2) value. The settings provided in this procedure
measure the period from falling edge to falling edge, which is not a valid measurement.
Select the clocks to be tested.
4
a Select the clock field to be tested and then select the clock edges as indicated in the
table.
Clocks
J↑ + K↑ + L↑ + M↑
b Connect the rising edge clocks to the pulse generator channel 1 output.
c Select Done to exit the Master Clock menu.
To test the multiple-clock, multiple-edge, state acquisition (logic analyzer)
Testing Performance
3–37