Select Exit to exit the Board Verification Test. In the Analy PV menu, select
14
Acquisition IC Verification, then select Communication Test.
Select Run, then select Single.
15
To run a test continuously, select Repetitive. Select Stop to halt a repetitive test.
For a Single run, the test runs one time and the screen shows the results.
If you have an Agilent 1670G-series Logic Analyzer without an oscilloscope or pattern
generator, go to step 19. If you have an Agilent 1670G-series Logic Analyzer with a pattern
generator, go to step 18. If you have an Agilent 1670G-series Logic Analyzer with an
oscilloscope, continue with step 15.
Select Analy PV, then select Scope PV in the pop up menu. Select Functional Tests.
16
Troubleshooting
To run the self-tests
5–17