c On the oscilloscope, select [Shift] ∆ Time, then select [Enter] to display the setup time
(∆ Time(1)-(2)).
d Adjust the pulse generator channel 1 Delay until the pulses are aligned according the
the setup time of the setup/hold combination selected, +0.0 ps or -100 ps.
Disregard the oscilloscope Period(2) value. The settings provided in this procedure
measure the period from falling edge to falling edge, which is not a valid measurement.
Select the clock to be tested.
4
a In the Master Clock menu, select the clock field to be tested, then select the clock
edge as indicated in the table. The first time through this test, use the top clock and
edge in the following table.
Clocks
J↑
K↑
L↑
M↑
b Connect the clock to be tested to the pulse generator channel 1 output.
c Select Done to exit the Master Clock menu.
To test the single-clock, single-edge, state acquisition (logic analyzer)
Testing Performance
3–25