Connect The Logic Analyzer - Agilent Technologies 1670G Series Service Manual

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

Testing Performance
To test the multiple-clock, multiple-edge, state acquisition (logic analyzer)

Connect the logic analyzer

Using the 6-by-2 test connectors, connect the first combination of logic analyzer
1
clock and data channels listed in one of the following tables to the pulse generator.
If you are testing an Agilent 1670G or Agilent 1671G, you will repeat this test for the second
combination.
Using SMA cables, connect channel 1, channel 2, and trigger of the oscilloscope to
2
the pulse generator.
Connect the Agilent 1670G or Agilent 1671G Logic Analyzer to the Pulse Generator
Testing Combinations
1
2
3–32
Connect to
Connect to
8133A
8133A Channel 2 Output
Channel 2 Output
Pod 1, channel 3
Pod 2, channel 3
Pod 3, channel 3
Pod 4, channel 3
Pod 5, channel 3
Pod 6, channel 3
Pod 7, channel 3
Pod 8, channel 3 *
Pod 1, channel 11
Pod 2, channel 11
Pod 3, channel 11
Pod 4, channel 11
Pod 5, channel 11
Pod 6, channel 11
Pod 7, channel 11
Pod 8, channel 11 *
Connect to
8133A Channel 1 Output
J-clock
K-clock
L-clock
M-clock
J-clock
K-clock
L-clock
M-clock
*Agilent 1670G only.