To test the trigger sensitivity (oscilloscope)
Testing the trigger sensitivity verifies the performance of the following specifications:
•
DC to 50 MHz: 0.063 x full scale (0.25 division)
•
50 MHz to 500 MHz: 0.125 x full scale (0.5 division)
Equipment Required
Equipment
Critical Specifications
Signal Generator
50 MHz and 225 MHz, 30 - 80 mV RMS output
Cable
BNC 48-inch
Adapter
Type N (m) to BNC (f)
Set up the equipment
Turn on the equipment required and the logic analyzer. Let them warm up for 30
minutes if you have not already done so.
Recommended
Model/Part
8656B Opt. 001
8120-1840
1250-0780
3–81