Enable the pulse generator channel 1 output (with the LED off).
4
Set up the Waveform menu.
5
a Press Run to fill the acquisition memory so the markers can be configured.
b Press the Waveform key.
c Move the cursor to the sec/Div field, then use the knob to dial in 100 ns.
d Select the Markers Off field, then select Pattern.
e Select the Specify Patterns field. Select X entering 1 and O entering 1.
f Select Done to exit the Specify Patterns menu.
g Move the cursor to the X-pat field. Type 1, then press Done.
h Move the cursor to the O-pat field. Type 16384, then press Done.
i Select O-pat from Trigger field, and then select from X marker.
To test the time interval accuracy (logic analyzer)