Agilent Technologies 1670G Series Service Manual page 80

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

Testing Performance
To test the time interval accuracy (logic analyzer)
Set up the Format menu.
2
a Press the Format key. Select Timing Acquisition Mode, then select Full Channel
250 MHz.
b Select the field to the right of the Pod A1 field, then select ECL.
c Select the field showing the channel assignments for Pod A1. Deactivate all channels
by pressing the Clear entry key. Using the arrow keys, move the selector to Channel 0.
Press the Select key to put an asterisk in the channel position, activating the channel,
then press the Done key.
Set up the Trigger menu.
3
a Select Acquisition Control. Select the Acquisition Mode field, and the acquisition mode
should toggle to Manual.
b Select Trigger Position. In the Pop-up menu, select Start.
c Select the Sample Period field. In the numeric pop-up menu, select 8 ns.
d Select the Memory Length field, and then use the RPG to select 65536.
e Press the Done key to exit the Acquisition Control menu.
3–56