Set up the logic analyzer

Set up the Configuration menu.
1
a Press the Config key. At the pop up menu, select Scope Channel.
b Select the Input field, then select C1.
c Move the cursor to the Probe field, press Select, then use the RPG knob to dial in 1:1.
d Move the cursor to the V/Div field and press Select. Type 80 on the front-panel
keyboard, then use the left and right control keys to select mV. Press the Select key.
e Move the cursor to the Offset field. Set the offset to 0 by typing 0, then pressing the
Select key.
f Select the Coupling field, then select 50Ω / DC.
g Move the cursor to the s/Div field, press Select, then use the RPG knob to dial in
200 ns.
Set up the Display menu.
2
a Press the Display key.
b Select the Mode field, then select Average.
c Move the cursor to the Average # field. Type 32 on the front-panel keyboard, then
press Done.
d Select the Grid field and set it to On.
e In the Waveform selection menu, delete channel 2. If channel 1 is not inserted, insert
channel 1.
Testing Performance
To test the bandwidth (oscilloscope)
3–73