Agilent Technologies 1670G Series Service Manual page 159

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

To verify pattern output (pattern generator option only)
Equipment Required
Equipment
Oscilloscope
Probe
Output Data Pod
Connect one of the 10460-series data pods to the end of the pattern generator Pod 1
1
cable.
Touch Output Patterns. In the pop-up menu, touch Checkerboard Pattern.
2
Using an oscilloscope, verify the existence of logic-level transitions by touching the
3
oscilloscope probe to each channel of Data Pod 1 and doing an Autoscale. The signal
levels that appear on the oscilloscope display should correspond with the logic
levels represented by the 10460-series pod being used.
To verify pattern output (pattern generator option only)
Critical Specification
≥ 500 MHz Bandwidth
500 MHz Bandwidth
no substitute
Troubleshooting
Recommended Model/Part
54522A
10441A
10460A - series
5–33