Testing Performance
To test the time interval accuracy (logic analyzer)

Acquire the data

Press the blue key, then press the Run key to select Run-Repetitive. Allow the logic
1
analyzer to acquire data for at least 1 minute. Observe the X to O time field and
ensure the X marker to O marker time is between 409.56 and 409.64 µs during the
test.
Record the results in the performance test record.