Testing Performance
To test the time interval accuracy (logic analyzer)
Set up the function generator according to the following table.
3
Function Generator Setup
Freq: 40.000 00 MHz
Amptd: 1.00 V
Modulation: Off
Connect the logic analyzer
Using a 6-by-2 test connector, connect channel 0 of Pod A1 to the pulse generator
1
channel 1 output.
Using the SMA cable and the BNC adapter, connect the External Input of the pulse
2
generator to the Main Signal of the function generator.
3–54