Agilent Technologies 1670G Series Service Manual page 67

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

Set up the Format menu.
2
a Press the Format key.
b Select the field to the right of each pod field, then select ECL. The screen does not
show all pod fields at one time. Use the knob to access pod fields not shown on the
screen.
Set up the Trigger menu.
3
a Press the Trigger key. Select Modify Trigger, then select Clear Trigger, then select All
in the pop-up menu.
b Select Count Off. Press Select again, then select Time in the pop-up menu. Select
Done to exit the menu.
c Select Acquisition Control. If the Acquisition Control is "Manual," use the cursor keys
to move the cursor to the Acquisition Mode field. Press the "Select" key to toggle this
field to "Automatic."
d In the Acquisition Mode menu, use the knob to set the Memory Length to 4096.
e Select the field labeled 1 under the State Sequence Levels. Select the field labeled
"anystate," then select "no state." Select Done to exit the State Sequence Levels menu.
To test the single-clock, multiple-edge, state acquisition (logic analyzer)
Testing Performance
3–43