Offset Frequency
Test limit [dB]
[Hz]
± 1.5
+7.5 M
CW Frequency: 3 GHz
Offset Frequency
Test limit [dB]
[Hz]
± 1.5
-7.5 M
± 1.5
-6.5 M
± 1.5
-5.5 M
± 1.5
-4.5 M
± 1.5
-3.5 M
± 1.5
-2.5 M
± 1.5
-1.5 M
± 1.5
-0.5 M
± 1.5
+0.5 M
± 1.5
+1.5 M
± 1.5
+2.5 M
± 1.5
+3.5 M
± 1.5
+4.5 M
± 1.5
+5.5 M
± 1.5
+6.5 M
± 1.5
+7.5 M
DC Power Voltage Test
DC POWER
Test limit [mV]
Voltage [V]
± 2.0
0
± 4.0
1
± 6.0
2
± 12.0
5
± 22.0
10
± 26.0
12
Chapter 2
E5052B Performance Test Record
Test result
Measurement
[dB]
uncertainty [dB]
± 0.287
Test result
Measurement
[dB]
uncertainty [dB]
± 0.314
± 0.361
± 0.308
± 0.360
± 0.368
± 0.325
± 0.333
± 0.370
± 0.320
± 0.328
± 0.337
± 0.373
± 0.339
± 0.330
± 0.377
± 0.318
Test result
Measurement
[mV]
uncertainty [mV]
± 0.140
± 0.159
± 0.181
± 0.243
± 0.366
± 0.429
Performance Tests
TAR
Status
6.05
PASS
TAR
Status
5.51
PASS
4.76
PASS
5.63
PASS
4.78
PASS
4.67
PASS
5.31
PASS
5.19
PASS
4.64
PASS
5.40
PASS
5.26
PASS
5.12
PASS
4.60
PASS
5.08
PASS
5.23
PASS
4.55
PASS
5.43
PASS
TAR
Status
14.28
PASS
25.15
PASS
33.14
PASS
49.38
PASS
60.10
PASS
60.60
PASS