Replacement Procedure

Front BNC Connector Replacement

Front BNC Connector Replacement

Tools Required

TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Cutting plier
Open-end wrench, 9/16 inch (set to 3.39 N-m / 30.0 lb-in)
Open-end torque wrench, 5/16 inch (set to 1.02 N-m / 9.0 lb-in) (for reconnecting SMA
connector)

Removal Procedure

Refer to
Figure 6-30
Step 1. Remove the outer cover as described in
Step 2. Remove the front panel as described
Step 3. Remove the DC Source board as described
page
170.
Step 4. Cut the cable tie (item 1) which ties cable.
Step 5. Remove the two TORX T10 screws (item 2) fastening the bracket BNC.
Step 6. Remove the three cable assembly (item 3) fastening the bracket BNC.
Step 7. If the diode is to be replaced, unsolder the diode (item 4).
Step 8. Remove a 9/16 inch nut (item 5) fastening the angle.
Step 9. Remove the three RF connector (item 6) fastening the angle.
192
for this procedure.
"Outer Cover Removal" on page
in"Front Panel Removal" on page
in"A2 DC Source Board Replacement" on
136.
138.