Replacement Procedure

Front USB PCA Board Replacement

Front USB PCA Board Replacement

Tools Required

TORX screwdriver, T10, T15, and T20
Open-end torque wrench, 5/16 inch (set to 1.02 N-m / 9.0 lb-in) (for reconnecting SMA
connector)
Flat edge screwdriver

Removal Procedure

Refer to
Figure 6-28
Step 1. Remove the outer cover as described in
Step 2. Remove the front panel as described in
Step 3. Remove the three TORX T15 screws (item 1).
Step 4. Disconnect the USB cables (item 2) from the front USB PCA.
Step 5. Remove the two TORX T10 screws (item 3) fastening the bracket BNC.