E5052B Performance Test Calculation Sheet; Introduction; Phase Noise Accuracy Test - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

E5052B Performance Test Calculation Sheet

Introduction

This section contains calculation sheets for each performance test that requires additional
calculations to determine the final test result.
Use the calculation sheet in this section as an aid for recording raw measurement data and
calculating the performance test results.
Calculation sheet entries are provided only for performance tests in which calculations are
required to obtain the test results.

Phase Noise Accuracy Test

Table 2-2
Sheet for Reference Frequency (10 MHz)
Offset
E5052B
Frequency
reading
[Hz]
[dBm]
E5052B
-1 M
-1 k
-100
100
1 k
1 M
Chapter 2
E5052B Performance Test Calculation Sheet
Power Meter reading
[dBm]
Setp Att [dB]
SG1
SG2
Cal value of
Test Result
[dB]
SATT
Result
Performance Tests
Test Result
equation
Result =
E5052B
Reading - (SG1 -
(SG2 - SATT))
27