Performance Tests
E5052B Performance Test Record
@ Frequency Band: 250 MHz - 7 GHz, Power Level: +20 dBm
Frequency
Frequency
[Hz]
Resolution [Hz]
301 M
64 k
301 M
1 k
301 M
10
7 G
64 k
7 G
1 k
7 G
10
Power Measurement Accuracy Test
@ -20 dBm
Frequency
Test limit [dB]
[Hz]
± 1.0
30 M
± 1.0
1 G
± 1.0
2.0125 G
± 1.0
3.0125 G
± 1.0
4.0125 G
± 1.0
5.0125 G
± 1.0
6.0125 G
± 1.0
7 G
@ -10 dBm
Frequency
Test limit [dB]
[Hz]
± 1.0
30 M
± 1.0
1 G
± 1.0
2.0125 G
± 1.0
3.0125 G
± 1.0
4.0125 G
± 1.0
5.0125 G
34
Test Limit
Test result
[Hz]
[Hz]
± 64 k
± 1 k
± 10
± 64 k
± 1 k
± 10
Test result
Measurement
[dB]
uncertainty [dB]
± 0.188
± 0.186
± 0.190
± 0.191
± 0.292
± 0.305
± 0.299
± 0.299
Test result
Measurement
[dB]
uncertainty [dB]
± 0.206
± 0.161
± 0.183
± 0.196
± 0.288
± 0.286
Measurement
TAR
uncertainty [Hz]
± 0.94 k
68.0
± 1.91
523.5
± 0.26
38.4
± 4.41 k
14.5
± 18.41
54.3
± 0.42
23.8
TAR
5.88
5.93
5.79
5.78
3.73
3.56
3.65
3.65
TAR
5.33
6.88
6.03
5.62
3.78
3.81
Status
PASS
PASS
PASS
PASS
PASS
PASS
Status
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
Status
PASS
PASS
PASS
PASS
PASS
PASS
Chapter 2