Power Measurement Accuracy Test - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

Performance Tests
E5052B Performance Test Record
@ Frequency Band: 250 MHz - 7 GHz, Power Level: +20 dBm
Frequency
Frequency
[Hz]
Resolution [Hz]
301 M
64 k
301 M
1 k
301 M
10
7 G
64 k
7 G
1 k
7 G
10

Power Measurement Accuracy Test

@ -20 dBm
Frequency
Test limit [dB]
[Hz]
± 1.0
30 M
± 1.0
1 G
± 1.0
2.0125 G
± 1.0
3.0125 G
± 1.0
4.0125 G
± 1.0
5.0125 G
± 1.0
6.0125 G
± 1.0
7 G
@ -10 dBm
Frequency
Test limit [dB]
[Hz]
± 1.0
30 M
± 1.0
1 G
± 1.0
2.0125 G
± 1.0
3.0125 G
± 1.0
4.0125 G
± 1.0
5.0125 G
34
Test Limit
Test result
[Hz]
[Hz]
± 64 k
± 1 k
± 10
± 64 k
± 1 k
± 10
Test result
Measurement
[dB]
uncertainty [dB]
± 0.188
± 0.186
± 0.190
± 0.191
± 0.292
± 0.305
± 0.299
± 0.299
Test result
Measurement
[dB]
uncertainty [dB]
± 0.206
± 0.161
± 0.183
± 0.196
± 0.288
± 0.286
Measurement
TAR
uncertainty [Hz]
± 0.94 k
68.0
± 1.91
523.5
± 0.26
38.4
± 4.41 k
14.5
± 18.41
54.3
± 0.42
23.8
TAR
5.88
5.93
5.79
5.78
3.73
3.56
3.65
3.65
TAR
5.33
6.88
6.03
5.62
3.78
3.81
Status
PASS
PASS
PASS
PASS
PASS
PASS
Status
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
Status
PASS
PASS
PASS
PASS
PASS
PASS
Chapter 2

Advertisement

Table of Contents
loading

Table of Contents