A25 24 Bit I/O Board Replacement; Tools Required; Removal Procedure - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

Replacement Procedure

A25 24 bit I/O Board Replacement

A25 24 bit I/O Board Replacement

Tools Required

TORX screwdriver, T10, T15, and T20

Removal Procedure

Refer to
Figure 6-8
Step 1. Remove the outer cover as described in
Step 2. Disconnect the flat cable (item 1) from the 24 bit I/O board.
Step 3. Remove the two TORX T10 screws (item 2) fastening the 24 bit I/O board.
Step 4. Slide the 24 bit I/O board.
150
for this procedure.
"Outer Cover Removal" on page
136.