Replacement Procedure

Power Supply Assembly Replacement

Power Supply Assembly Replacement

Tools Required

TORX screwdriver, T10, T15, and T20

Removal Procedure

Refer to
Figure 6-6
Step 1. Remove the outer cover as described in
Step 2. Remove the three TORX T10 screws (item 1) from the rear panel.
Step 3. Disconnect the power supply cables (item 2).
Step 4. Remove the two TORX T20 screws (item 3) fastening the plate for power supply.
Step 5. Lift the power supply assembly toward front side.