Performance Test Failure Troubleshooting; Recommended Adjustment For Performance Test Failure - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

Troubleshooting

Performance test failure troubleshooting

Performance test failure troubleshooting
This section describes the adjustment and troubleshooting procedures used when the
E5052B fails the performance tests. If the performance of the instrument is critical for the
test limits and seems to be adjustable, perform first the adjustment(s) related to the failed
test. When the test result is far from the tolerance of the test or the performance is not
adjustable, isolate the faulty assembly in accordance with the "Performance tests failure
troubleshooting procedure".

Recommended adjustment for performance test failure

Table 4-3
adjustment program corresponding to the recommended adjustment and perform the
adjustment. Note that adjustment items for theA3 or A4/5 are the same as the A6.
When RF IN Port VSWR test fails, check first whether the connections of the RF
semi-rigid cables are tight or loose. Also check for possible disconnection (impairment) of
the cables and connectors.
Table 4-3
Recommended adjustment for performance test failure
Test
Num.
1
2
3
4
5
6
7
8
78
shows the recommended adjustments when the performance test fails. Select the
First failed test
RF IN Port VSWR Test
Frequency Measurement Test
Power Measurement Accuracy Test
Phase Noise Accuracy Test
SSB Phase Noise Sensitivity Test
SA Relative Level Accuracy Test
DC Power Voltage Test
DC Power Current Measurement Test
Recommended adjustment
A1
A2
A3
A4/5
Chapter 4
A6

Advertisement

Table of Contents
loading

Table of Contents