Firmware Update

Firmware Update
Firmware Update

Required Equipment

NOTE
The free space of 30 MB or more is nessessary for the USB mass strage device.

How to make E5052B Firmware Installation Device

The following is the procedure to make the E5052B Firmware Installtion Device.
Step 1. Copy the following files from the Keysight Technologies web site to the USB mass storage
device.
The latest firmware can be downloaded from the following site.

How to update the E5052B Firmware

It is possible to update the E5052B firmware, but not the following software by this
procedure.
NOTE
To execute the firmware update, you must first disable a USB Mass Storage Divice. For
more on how to disable a USB Mass Storage Divice, refer to Online Help.
The following is the procedure to update the E5052B firmware.
Step 1. Download the latest firmware from the download site. It is prepared as execution file
(E5052B_xx.xx.exe).
Step 2. Run the "E5052B_xx.xx.exe" to extract the E5052.msi and.setup.vbs. Then, save these
files to USB mass storage devie.
Step 3. Connect the mouse and keyboard to the E5052B rear panel. Then turn the E5052B on.
Step 4. Connect the USB mass storage divice into the front USB port of the E5052B.
Step 5. Update the E5052B firmware.
1. Press
236
E5052B (with keyboard and mouse).
USB memory.
E5052.msi
Setup.vbs
http://www.keysight.com/find/ssa/
OS (Operating System)
VBA (Visual Basic for Application)
Calibration Constand Data
key.