Adjustment
Required Adjustment after Replacing Assembly

PM Flatness Adjustment

The purpose of this procedure is to adjust the PM flatness of the A1 freq/power detector
module.
Required equipment for the PM flatness adjustment
Description
Signal Generator
Power Splitter
Power Sensor
Power Meter
N(m)-N(m) Cable
N(m)-N(m) Adapter

Receiver RF Flatness Adjustment

The purpose of this procedure is to adjust the RF flatness of the A6 phase detector board.
Required equipment for the receiver RF flatness adjustment
Description
Signal Generator
50 ohm (BNC) Termination
Power Sensor
Power Meter
Power Splitter
N(m)-N(m) Cable
N(m)-N(m) Adapter
48
Recommended Model
Keysight E8257D with Opt. UNX, UNT, 1EA
without Opt. UNW, 1EH
Keysight 11667A
Keysight E9304A Opt. H18 or Keysight E9304A
Opt. H19
Keysight E4419A/B
Keysight N6314A (p/n 8120-8862)
Keysight p/n 1250-1475
Recommended Model
Keysight E8257D with Opt. UNX, UNT, 1EA
without Opt. UNW, 1EH
Keysight 11593A
Keysight E9304A Opt. H18 or Keysight E9304A
Opt. H19
Keysight E4419A/B
Keysight 11667A
Keysight N6314A (p/n 8120-8862)
Keysight p/n 1250-1475