Replacement Procedure

Outer Cover Removal

Outer Cover Removal

Tools Required

TORX screwdriver, T10,T15 and T20

Procedure

Refer to
Figure 6-1
Step 1. Disconnect the power cable from the E5052B.
Step 2. Remove the four bottom feet (item 1).
Step 3. Remove the four TORX T10 screws (item 2) from the bottom.
Step 4. Remove the four TORX T20 screws (item 3) fastening the side strap handles.
Step 5. Remove the four TORX T15 screws (item 4) fastening the four rear foot
Step 6. Remove the four TORX T15 screws (item 5) fastening the cover.
Step 7. Slide the outer cover and remove it carefully.
136
for this procedure.
Chapter 6