Pm Offset Adjustment - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

Adjustment
Required Adjustment after Replacing Assembly

PM Offset Adjustment

The purpose of this procedure is to adjust the PM offset of the A1 freq/power detector
module.
Required equipment for the PM offset adjustment
Description
Signal Generator
Power Amplifier (35dB)
Step Attenuator
Attenuator (3dB)
Blocking Capacitor
Low Pass Filter
Power Splitter
Power Sensor
Power Meter
N(m)-N(m) Cable
N(m)-N(m) Adapter
N(f)-SMA(m) Adapter
46
Recommended Model
Keysight E8257D with Opt. UNX, UNT, 1EA
without Opt. UNW, 1EH
R & K A3000-2H-R
Keysight 8496H
Keysight 8491A Opt. 003
Keysight 11742A
Mini-Circuits SLP-70 (70MHz)
Keysight 11667A
Keysight E9304A Opt. H18 or Keysight E9304A
Opt. H19
Keysight E4419A/B
Keysight N6314A (p/n 8120-8862)
Keysight p/n 1250-1475
Keysight p/n 1250-1562
Chapter 3