Dc Current Measurement Test - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

Performance Tests
E5052B Performance Test Record
DC POWER
Test limit [mV]
Voltage [V]
16

DC Current Measurement Test

DC Current [mA]
Test limit [μA]
1
2
5
10
20
50
80
40
Test result
[mV]
± 34.0
Test result
[μA]
± 162.0
± 164.0
± 170.0
± 180.0
± 200.0
± 260.0
± 320.0
Measurement
TAR
uncertainty [mV]
± 0.533
63.78
Measurement
TAR
uncertainty [μA]
± 4.146
39.07
± 5.475
29.95
± 4.688
36.26
± 31.619
± 7.658
26.11
± 17.210
15.10
± 79.038
Status
PASS
Status
PASS
PASS
PASS
5.69
PASS
PASS
PASS
4.04
PASS