Oven Ch2 Assembly Replacement; Tools Required; Removal Procedure - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

Replacement Procedure

Oven CH2 Assembly Replacement

Oven CH2 Assembly Replacement

Tools Required

TORX screwdriver, T10, T15, and T20

Removal Procedure

Refer to
Figure 6-12
Step 1. Remove the outer cover as described in
Step 2. Remove the four TORX T10 screws (item 1) fastening the oven ch2 Assembly.
Step 3. Disconnect the RF cable (item 2) from the oven ch2 Assembly.
Step 4. Disconnect the cable (item 3) from the connector.
Step 5. Release the cable (item 4) from the cable clamp.
Step 6. Disconnect the connector (item 5) from the oven ch2 Assembly.
Step 7. Lift the oven ch2 Assembly (item 6).
Step 8. Remove the three TORX T10 screws (item 7) fastening the C-bracket.
Step 9. Remove the oven ch2 (item 8).
158
for this procedure.
"Outer Cover Removal" on page
136.
Chapter 6