Replacement Procedure

Oven CH1 Assembly Replacement

Oven CH1 Assembly Replacement

Tools Required

TORX screwdriver, T10, T15, and T20

Removal Procedure

Refer to
Figure 6-11
Step 1. Remove the outer cover as described in
Step 2. Remove the three TORX T10 screws (item 1) fastening the oven ch1 assembly.
Step 3. Disconnect the RF cable (item 2) from the oven ch1 assembly.
Step 4. Disconnect the cable (item 3) from the power distribution board.
Step 5. Lift the oven ch1 assembly (item 4).
Step 6. Remove the two TORX T10 screws (item 5) fastening the C-bracket.
Step 7. Remove the three TORX T10 screws (item 6) fastening the C-bracket.
Step 8. Remove the oven ch1 (item 7).
Step 9. Remeve the L-bracket (item 8).