Phase Noise Accuracy Test - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

Performance Tests
E5052B Performance Test Record
Frequency
Test limit [dB]
[Hz]
± 1.0
1 G
± 1.0
2.0125 G
± 1.0
3.0125 G
± 1.0
4.0125 G
± 1.0
5.0125 G
± 1.0
6.0125 G
± 1.0
7 G

Phase Noise Accuracy Test

@ Frequency : 1 GHz
Offset
Test limit [dB]
Frequency [Hz]
- 100 M
- 40 M
- 10 M
- 1 M
- 1 k
- 100
100
1 k
1 M
10 M
40 M
100 M
@ Frequency : 10 MHz
Offset
Test limit [dB]
Frequency [Hz]
- 1 M
- 1 k
36
Test result
[dB]
Test result
[dB]
± 3.0
± 2.0
± 2.0
± 2.0
± 2.0
± 3.0
± 3.0
± 2.0
± 2.0
± 2.0
± 2.0
± 3.0
Test result
[dB]
± 2.0
± 2.0
Measurement
uncertainty [dB]
± 0.158
± 0.150
± 0.170
± 0.282
± 0.275
± 0.273
± 0.276
Measurement
uncertainty [dB]
± 0.231
± 0.236
± 0.242
± 0.238
± 0.237
± 0.291
± 0.282
± 0.240
± 0.246
± 0.234
± 0.236
± 0.229
Measurement
uncertainty [dB]
± 0.252
± 0.257
TAR
Status
3.30
PASS
3.47
PASS
6.49
PASS
3.86
PASS
3.96
PASS
3.99
PASS
3.95
PASS
TAR
Status
18.22
PASS
10.46
PASS
10.22
PASS
10.36
PASS
10.42
PASS
14.35
PASS
14.85
PASS
10.30
PASS
10.06
PASS
10.59
PASS
10.46
PASS
18.39
PASS
TAR
Status
9.79
PASS
9.60
PASS
Chapter 2

Advertisement

Table of Contents
loading

Table of Contents