Replacement Procedure

Front Panel Removal

Front Panel Removal

Tools required

Procedure

Refer to
Step 1. Remove the outer cover as described in
Step 2. Disconnect the two semi-rigid cables (item 1) from the front panel.
NOTE
When reconnecting the semi-rigid cables, make sure that the cables are connected to the
same connectors from which they previously came.
Step 3. Remove the top trim (item 2) using a flat edge driver.
Step 4. Remove the seven TORX T15 screws (item 3 ) fastening front panel.
Step 5. Release the flat cables from the cable clamps (item 4).
Step 6. Gradually push the front panel assembly towards the outside.
Step 7. Disconnect the cable (red and black) ( item 5) from the front panel.
Step 8. Disconnect the cable (item 6) from the front panel.
138
TORX screwdriver T10, T15 and T20
Flat edge screwdriver
Open torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Figure 6-2
for this procedure.
"Outer Cover Removal" on page
136.