Table 1-4 Test Record For Offset Accuracy Verification - Keysight 8990B Service Manual

Peak power analyzer
Hide thumbs Also See for 8990B:
Table of Contents

Advertisement

12 Calculate the V
13 Repeat
14 Repeat
15 Repeat
Table 1-4
Test record for offset accuracy verification
Vertical scale
Output
(V/d iv)
vol tage
Video input channel 2 (50 Ω)
0.01
±120 mV
0.02
±240 mV
0.05
±600 mV
0.1
±1.2 V
0.2
±2.4 V
0.5
±4.0 V
1
±4.0 V
Video input channel 2 (1 MΩ)
0.01
±5 V
0.02
±10 V
0.05
±10 V
0.1
±20 V
0.2
±20 V
0.5
±20 V
1
±100 V
2
±100 V
5
±100 V
Keysight 8990B Service Guide
error+
record the value in
Table
V
= V
– V
error
DMM
scope
step 5
to
step 12
step 5
to
step 13
step 3
to
step 14
V
V
DMM+
scope+
Performance Verification and Adjustments
and V
respectively using the following equation and
error–
1-4.
for the rest of the vertical scale setting in
with the input impedance set to 1 MΩ.
for video input channel 3.
V
V
error+
DMM–
V
V
V
scope–
error–
zero-error
1
Table
1-4.
V
zero-error
limit (±)
1.8 mV
2.6 mV
5.0 mV
9.0 mV
17.0 mV
41.0 mV
81.0 mV
1.8 mV
2.6 mV
5.0 mV
9.0 mV
17.0 mV
41.0 mV
81.0 mV
161.0 mV
401.0 mV
41

Advertisement

Table of Contents
loading

Table of Contents