Performance Verification
Voltage tests
Test each output channel individually. Refer to the appropriate test record form.
NOTE
Source voltage mode programming and readback accuracy
This test verifies that the voltage programming and readback through the USB are
within published specifications.
Test settings
Table 2-2
Test settings for source voltage mode programming and readback accuracy
Function
DC voltage
[a] Use the highest resolution available on the DMM.
Keysight U2722A/U2723A Service Guide
Performance Verification and Adjustment
[a]
Keysight 3458A settings
Range
1 V (for U2722A/U2723A drives ≤ 1 V)
10 V (for U2722A/U2723A drives ≤ 10 V)
2