To Update the Firmware
If a firmware update is available, perform the following procedure to update the
firmware of the M9602A/M9603A modules.
1. Click Start menu > Keysight M960x Source Measure Unit > M960x SFP. The
2. In the list on the window, highlight the M9602A/M9603A modules to connect,
3. Confirm that the Access indicator on the M9602A/M9603A front panel turns
4. Select Firmware Update... from the Utilities menu. The Firmware Update
5. Select the latest version of the firmware, if it is available.
6. Click Update at the bottom of the Firmware Update dialog box to start the
7. Close the M960x SFP.
8. Restart your system; Power down the controller, then the chassis, and power
Keysight M9602A/M9603A Startup Guide, Edition 3
Connect to Instrument dialog box will open.
and click Connect to launch the Keysight M960x PXIe Source Measure Unit
Soft Front Panel (the M960x SFP).
green. Refer to
Figure 1-1
dialog box will open.
If the selection menu displays "No change", there are no firmware updates.
firmware update.
When the firmware is updated successfully, the message appears in the
Firmware Update dialog box.
up the chassis, then the controller.
Step 5: Verify Operation of the Module
and
Table 1-3
for LED status.